GaugeType
B-A Gauge
Pressure Range
1.0 x E-1 ~ 1.0 x E-9 mbar
Main AC
100~240V 50/60H
Power Consumption
28W (measurement mode with 5m cable)
Filament
Yttria coated Iridium x2 (Auto Switching)
Emission current
42mA
2.5mA
Tough Mode
300ºC | 500ºC | 700ºC | OFF
Tough Mode Heating Part
Grid & Collector
Degas Mode
Grid & Collector: 900ºC (2min)
Operating Temperature
0 ~ 40ºC
Inputs
Emission ON/OFF
Tough Mode ON/OFF
Degas ON/OFF
Outputs
Analog 0~10V x 2
Setpoint x 2, PhotoMOS relay N.O. Form A
Alam x 1, PhotoMOS relay N.O. Form A
Dimensions
1/4 Rack(3U) x 230mm Depth
Weight
Controller: 1.2kg
I/O Connectors
Dsub15 / USB
Flange
KF25, CF40
Communication
RS232C, USB
Remarks
Sensor Cable 5m

B-A Tough Gauge© had been developed based on the patented Tough Mode© technology by Ampere Inc. in Japan.
The Tough Mode© can heat up the grid and the collector while pressure measurement in order to prevent contaminants in process environment from being deposited on the grid and the collector. In conventional B-A Gauge, the deposited contaminants shorten the lifetime of the gauge and cause inaccurate measurement.
The B-A Tough Gauge© will give longer lifetime in the environment where corrosive or organic particles involved and will be able to keep the accuracy of the measurement all through your production cycle.

TG102S is a Byard-Albert ionisation gauge which can measure down to 10-9 mbar (hPa) with Tough Mode©. The external controller is rack mountable.

Part codes

TG Part CodeTG Part SpecificationsTG Part Contents
TG102S-001KF25, Viton sealSensor, Controller & Cable 5m
TG102S-002CF40, Metal gasketSensor, Controller & Cable 5m
TGH100-001KF25, Viton sealSensor onnly
TGH100-002CF40, Metal gasketSensor only
TGC102S-001Connection to a sensor with a cableController only
TG102SCBL-0055mCable